Class 1 Microbiological Safety Cabinets
LEEC Class 1 Microbiological Safety Cabinets
Designed to comply with BS EN 12469:2000
Features:
- Stainless steel contruction
- Digital display of air Velocity (m/sec)
- Digital Hours Run meter
- Audible & visual alarm system
- Externally mounted workspace light
- Two 13A splash proof electrical sockets
- Air tight steel night door
- Radius workspace corners
- Alarmed viewing window
- Easy to clean and disinfect
- Two sizes: 1m or 1.5m wide
- Touch operated controls
Class 1 definition: Safety cabinet with a front aperture through which the operator can carry out manipulations inside the cabinet and which is constructed so that the worker is protected and the escape of airborne particulate contamination generated within the cabinet is controlled by means of an inward airflow through the working front aperture and filtration of the exhaust air.
CONSTRUCTION
The new WL4E and WL4ES are made from stainless steel and is finished in white polyester powder coat paint for a durable, easy clean finish. The interior features a non-painted stainless steel floor with radiused corners to give a hard wearing surface which is easy to keep clean. The whole work area is evenly illuminated. The front edge has a raised arm rest to reduce operator fatigue. The steel night door is easily clamped into position, guaranteeing a gas tight seal without the need for taping, even during fumigation. A 6mm thick toughened glass window is hinged to allow easy access to the interior for cleaning, pre-filter changing and placing bulky objects inside. The anti-blowback valve is fitted adjacent to the HEPA filter, and combines an airflow indicator and the 'airflow out of limits' warning system. An intuitive touch operated control panel features a digital readout of air velocity and also an 'hours run' meter.
AIR FILTRATION
The main HEPA filter is used at only 75% of its rated capacity (with NaCI penetration not exceeding 0.003% when tested to BS3928) so that it is never over stressed.
Air is drawn through the cabinet front, across the work area and through a pre-filter which greatly extends the useful working life of the main HEPA filter. Filtered air is drawn past the airflow indicator and through the installed ductwork to a safe discharge point by an exhaust fan.
INSTALLATION AND COMMISSIONING
Installation and commissioning must be carried out by qualified engineers. LEEC offers this service and also post installation maintenance / testing contracts.
WARRANTY
LEEC Class 1 Microbiological Safety Cabinets are covered by a 1 year warranty.

| SPECIFICATION |
||
| MODEL | DIMENSIONS (H x W x D) - mm | WORKING WIDTH (M) |
| WL4E | 1223 x 1128 x 660 | 1 metre (1 user) |
| WL4S | 1223 x 1638 x 660 | 1.5 metres (2 users) |